LED SHADOWLESS OPERATING LIGHT


LED SHADOWLESS OPERATING LIGHT (MV-2X700)
FEATURES
Color rendering index ≥ 90, near to real color as soon as possible
Three lighting modes
Standard Mode
Product characteristic
Using high brightness LED cold light source, the service life of LED light emitting secondary tube is more than 50,000 hours, light decay is low, energy saving, service life is long, and the Composition of ultraviolet and infrared ray is very low.
Stepless PWM dimming to satisfy brightness requirement for different surgery
High color temperature can improve the visual resolution of tissue, and low
color temperature can relieve the visual fatigue caused by long-term surgery
Three Lighting Mode suit for various operation with different lighting
Endoscopy Mode
Low Light Mode
Using digital dimming technology to meet the doctor’s requirements for different brightness, and stable color temperature.
The control board adopts the high quality constant current chip with constant current output, the current is more accurate, the quality is more reliable.
LED dot matrix series and arrangement, even if the individual LEDs are not lit, the overall brightness is rarely affected.
The spot can be continuously adjusted to meet the requirements of different doctors for the visual field of surgery.
The spot can be continuously adjusted to meet the requirements of different doctors for the visual field of surgery.
High color index, to meet the doctor’s blood vessels, internal organs and different tissue discrimination.
Low temperature rise, the temperature rise of the exposed surface is converted into the measurement of total irradiance. The total radiant illuminance Ee of an area 1m away from a single lamp cap does not exceed 1000W/m²
Lighting performance: with excellent shadowless rate, deep cavity lighting, light column depth.
MODEL
MV-2X700D
Main Parameter
700D
(@ Lamp Head)
700D
Work Temperature
5℃~40℃
Relative Humidity
Less than 80%
Work Voltage
AC 220V
Power Frequency
50Hz
Power Supply
AC 220V 0.6A
Operation Mode
Continuous Operation
Input Power
100VA
Ultraviolet Radiation
UV radiation below 400nm should not exceed 10W/m2
Center illumination Ec
40kLx-160kLx
Spot Diameter D10
180-280
Spot distribution(D50/D10)
>0.5
Residual illumination of single mask
≥50%
Double shade residual illumination
≥40%
Shadow Compensation
NO
Light Depth(mm)
≥600
Color Temperature
4000K≤Tc≤6500K
Color Rendering Index
Ra≥90
Total Radiation Measurement
≤1000W/m2
Install Height (cm)
280-320